Leica ASP300
5.1 Reagent handling
5.1.1 Fill/drain reagents (other than wax)
1. Manually changing the reagents
Remove the appropriate reagent bottle from the Leica ASP300 re-
agent cabinet, remove the bottle screw cap.
Drain/refill the reagent by pouring to/from a bulk container.
On the REAGENT STATUS screen, verify correct allocation and set
the reagent bottle as full.
2. Use the ‘Remote Fill / Drain’ features of the Leica ASP300
5. Operation
Connect the supplied remote fill/drain hose to
the remote drain connection (refer Fig. 11).
Important!
Ensure that the hose is pushed all of
the way onto the connection, until a
clear click is heard
Put the loose end of the remote fill/drain hose
into the desired container for the drain/fill
operation as desired.
Fig. 11